Intelect legend 2 coupling error
0
Fixing a “Coupling Error” on the Intelect Legend 2 – Learn about the common intelect legend 2 coupling error.

The Intelect Legend 2 differs from other units in how ultrasound gel should be applied. Applying gel directly to the sound head can sometimes trigger a “Coupling Error.”

Understanding the intelect legend 2 coupling error is crucial for effective use of the device.

To resolve the intelect legend 2 coupling error:

  1. Turn off the unit.
  2. Thoroughly clean the sound head, removing all ultrasound gel.
  3. Apply ultrasound gel to the patient’s skin only (not the sound head).
  4. Start/initiate the ultrasound therapy before placing the sound head on the patient.
  5. Once the treatment has started, then place the sound head onto the gelled treatment area.

Key point:
Always keep the sound head completely clean and apply gel only to the patient’s skin. Starting the treatment before contact prevent the coupling error

A complete manual for the intelect legend 2 can be found: HERE